医疗器械总结性可用性测试中使用错误的可检测性:测试环境的影响

Detectability of use errors in summative usability tests of medical devices: Impact of the test environment

Applied Ergonomics · 2024
被引 3
ABS 3

中文导读

研究了测试环境逼真度对医疗器械使用错误检测的影响,发现高逼真度不增加总体错误数,但影响特定错误的检测,建议优化环境前先分析各特征的作用。

Abstract

The regulations on summative usability evaluations of medical devices (MDs) emphasize that the test environment must have sufficient ecological validity for generalization to real-life use. Here, we examined the influence of environmental fidelity (a component of ecological validity) on the detectability of MD use errors. A total of 140 participants participated in a summative usability evaluation of an anaphylactic shock auto-injector device under either a high-fidelity condition or a condition acceptable from the manufacturer's perspective, lower-fidelity condition. The numbers of errors detected in each condition were compared by applying descriptive statistics and logistic and Poisson multivariate regressions. We found that the level of fidelity did not influence the overall number of use errors detected but did influence the detection of certain use errors. To optimize the test environment and increase the detection of use errors, each environmental feature's role in the test task should first be examined.

医疗器械可用性测试人机交互生态效度